Lab Test Details
SIM |
50205 |
CPT |
80299 |
Synonyms |
MTX |
Specimen Type |
Serum |
Container |
4.0 ml Gold-top serum separator |
Collection Instructions |
None |
Volume |
2 ml serum, minimum 0.5 ml |
Additional Information |
Specimens from patients who have received preparations of mouse monoclonal antibodies for diagnosis or therapy may contain human anti-mouse antibodies (HAMA). HAMA present in serum may interfere in this assay. These samples should not be assayed with the MTX assay. |
Patient Preparation |
none |
Specimen Processing Instructions |
Serum: Allow to clot. Centrifuge at 3380 rpm’s (±100) for a minimum of 10 minutes Centrifuge within 2 hours of collection. Protect from Light if not stored in a dark refrigerated. |
Transport Requirements |
Refrigerated |
Specimen Stability |
Room temperature: 8 hours Refrigerated: 24 hours |
Rejection Criteria |
If not centrifuged within 2 hours of collection. |
Reference Range |
NO NORMAL VALUES. Lowest reportable = less than 0.2 umol/L |
Critical Value |
none |
Test Includes |
Methotrexate |
Methodology |
Enzyme Immunoassay |
Testing Frequency |
Daily, Monday - Sunday |
|
|
Reviewed By |
Crystal M. Killian |
Reviewed Date |
3/18/2021 |
Supplemental |
|