Lab Test Details

Lab Test Details

SIM

LAB54

CPT

82330

Synonyms

Calcium Free, Serum ; Calcium Unbound, Serum; Calcium Filterable; Ca++, Ionized 

Specimen Type

Serum 

Container

Gold-top only

Collection Instructions

Cannot be shared with other specimens.

Volume

Volume: 1 ml. 
Min. volume: 1 ml

Additional Information

Evaluates non-bound calcium; if the tube is opened , the serum is exposed to air which causes a loss of carbon dioxide to the atmosphere. This changes the sample pH and can prevent lab from measuring the ionized calcium

Patient Preparation

 

Specimen Processing Instructions

DO NOT OPEN TUBE!  Place a label on tube that states DO NOT OPEN. Tube cannot be shared for other tests

Transport Requirements

Room temp or refrigerate; Do not freeze

Specimen Stability

14 days with unopened tube 

Rejection Criteria

Hemolysis; opened gel-barrier tube; gel-barrier not completely sealing the clot; plasma specimen; gross lipemia

Reference Range

Reference Range accompanies report

Critical Value

 

Test Includes

Ionized calcium 

Methodology

Ion-selective electrode (ISE) 

Testing Frequency

 

LCA

004804 

Reviewed By

Rhonda Crowe            

Reviewed Date

2/27/23

Supplemental

* Denotes Reference Lab Test